Thermocouple Lab Manual Artemis Ma PDF

Title Thermocouple Lab Manual Artemis Ma
Author Artemis Ma
Course Basic Mechanical Measurements Laboratory
Institution University of Minnesota, Twin Cities
Pages 4
File Size 280.8 KB
File Type PDF
Total Downloads 60
Total Views 134

Summary

Download Thermocouple Lab Manual Artemis Ma PDF


Description

UNIVERSITY OF MINNESOTA Mechanical Engineering ME 4031W

K-type Thermocouple Time Response Lab Manual Artemis Ma(ID: 4813358) Lab Section: 16 (TA: Manikandan Suresh) Specific Comments:

Examples of Effective Writing:

Suggestions for Improvement:

Additional Comments:

1. Introduction This document describes an outline of procedures to setup a thermocouple and a DAQ to measure temperature in Celsius, using a LabView virtual instrument. A thermocouple measures the temperature difference b

2. Hardware Guide & Set-Up Table 1. List of Equipment Materials/Equipment DAQ Digital Multimeter (DMM) Op-Amp Thermocouple Ice Water Bath

Type USB-6001 Keithley 2700 Gain of 100 K3

Uncertainty ± 6 mV [1] 0.0015% [2] – ± 2.2 °C [3] –

Figure 1: Schematic diagram of the thermocouple circuit[3] 1. Calibrating the Op-Amp: i Plug the op-amp to a power outlet and let it warm up for 15 minutes. ii Short the input sockets iii Connect the output sockets to the Keithley 2700 DMM. Adjust the potentiometer knob on the op-amp until the output reads close to zero [V]. 2. Setting up for measurement: i Connect the K-type thermocouple to the Op-Amp input with alligator clamp wires ii Set up the reference environments for the thermocouple. (For example, to measure room temperature, submerge the measuring tip of the thermocouple into the ice bath, leaving the reference ends in air) iii Connect the Op-Amp output to the DAQ (Vout- to Ai0+, and Vout- to the Ai0) iv Connect the DAQ to the Computer using the USB cable provided with the DAQ v Follow the LabView VI Guide

3. LabView VI Guide:

Figure 2: Screenshot of the LabView VI front panel 1. Create a csv file. This will be where the measurement data outputs to. 2. In the front panel of the VI, click on the folder icon and navigate to the csv file. 3. Input the DAQ measurement parameters into their respective boxes (“Rate [HZ]” as in rate of data collection, “number of samples” as in how many samples the DAQ would group into one data point) 4. Now you can click “run” on the top left to start measuring and recording of the temperature difference between the tip of thermocouple and the reference ends. 5. To correctly measure the temperature response, you can monitor the recordings in real-time with the waveform graph window. If measuring a stable temperature, the waveform may seem noisy. This is due to the autoscaling of the graph and not representative of the precision. 6. Press “stop” to stop recording

4. Troubleshooting Guide o If the temperature measurement remains around zero, the reference tips of the thermocouple that were submerged in ice water might be touching the metal container, therefore shorting the thermocouple. Adjust the set-up to resolve. o If the temperature reading is not zero when it should be, the op-amp might be producing an offset. Recalibrate the op-amp to resolve. o If the temperature reading appears to be the negative value of what it should be, switch the leads from the thermocouple going into the Op-Amp inputs to resolve.

References [1]

National Instruments. (2014). USB 6001 DAQ; Specifications. Retrieved from http://www.ni.com/pdf/manuals/374369a.pdf [2] Keithley. (2016). Digital Multimeter; Specifications. Retrieved from https://www.tek.com/datasheet/2700-multimeter-data-acquisition-switchsystems/integraseries-technical-data-sheet [3] Lab 1a LabView: Thermocouple Time Response...


Similar Free PDFs